ClearanceBrief
Recall Z-0460-2025 ยท FQO Table, Operating-Room, Ac-Powered

Getinge Usa Sales Inc

MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 710001B0

DateNov 14, 2024 (initiated Oct 3, 2024)
StatusOpen, Classified
ReasonError code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay
Root causeComponent design/selection
ActionMaquet GmbH, a subsidiary of Getinge (Maquet/Getinge) issued Urgent Medical Device Correction (FSCA 8010652-09/25/2024-002-C) letter on 03 October 2024, via FedEx delivery. Letter states reason for recall, health risk and action to take: Please examine your inventory immediately to determine if you have any of the MEERA Operating Tables with the Serial numbers listed in this notice. Note: The Product REF number (REF) and the Serial Number (SN) can be found on the product label. If the error code "50037" appears-(note below picture of the remote control displaying the Error Code), it will be
Quantity112 units
DistributionWorldwide distribution - US Nationwide and the countries of Australia, Austria, Bhutan, Brazil, Canada, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Ecuador, Egypt, France, Germany, Greece, India, Israel, Japan, Jordan, Korea, Republic Of Kuwait, "Macedonia, The Former Yugoslav Republic Of
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.