Recall Z-0278-2025 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Smith & Nephew Inc

smith&nephew LEIGON OXINIUM CONSTRAINED NONPOROUS FEMORAL COMPONENT, SIZE 4 LEFT, REF 71421164; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER

DateOct 30, 2024 (initiated Oct 2, 2024)
StatusOpen, Classified
ReasonComplaint received indicating that a package contained a LGN OX CONSTRAINED FEM 3 LT implant instead of a LGN OX CONSTRAINED FEM 4 LT as described on the product label.
Root causePackaging change control
ActionSmith & Nephew issued an "Urgent Medical Device Recall Notice" to its consignees on 10/02/2024 via email. The notice explained the issue, potential risk of use of the affected product, and requested the return of the affected devices. Distributors, sales reps, and district offices were directed to notify their customers. If you have any questions or concerns regarding this recall please contact FieldActions@smith nephew.com.
Quantity2 units
DistributionUS Distribution to state of Texas and OUS (foreign) to countries of: Spain, France and United Kingdom,.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K043440SMITH & NEPHEW, INC. REVISION KNEE SYSTEM
Smith & Nephew, Inc., Feb 18, 2005