Recall Z-0206-2025 ยท MHY Stimulator, Electrical, Implanted, For Parkinsonian Tremor

Medtronic Neuromodulation

Medtronic Percept RC Implantable Neurostimulators (INS), Model B35300.

DateOct 23, 2024 (initiated Sep 24, 2024)
StatusOpen, Classified
ReasonA limited number of Percept RC Implantable Neurostimulators (INS), Model B35300 will not be able to communicate with the A610 DBS Clinician Programmer Application upon use and an "Invalid Device" message will appear.
Root causeUnder Investigation by firm
ActionMedtronic neuromodulation issued an URGENT MEDICAL DEVICE RECALL notice to its consignees of 09/24/2024 via letter. The notice explained the issue, potential risk, and requested the affected devices in stock be returned. The notice is to be shared with those to whom the device was distributed. If the device has been implanted, the consignee was directed to contact their Medtronic Representative to determine if there are additional steps that may establish communication with the INS. On 07-Oct-24 and 08-Oct-24, a Medtronic representative successfully performed an update using a manufac
Quantity39 units
DistributionInternational distribution to the countries of Germany, UK, and Israel.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.