Recall Z-0292-2025 ยท CBK Ventilator, Continuous, Facility Use
Philips Respironics, Inc.
Trilogy Evo Universal Ventilator. Model Number DS2000X11B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.
| Date | Nov 13, 2024 (initiated Oct 7, 2024) |
| Status | Open, Classified |
| Reason | Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements. |
| Root cause | Device Design |
| Action | An URGENT Medical Device Correction notification was mailed to consignees on 10/07/2024. The notification provides instructions to users to prevent risks for patients when using in-line nebulizers and a set FiO2. Consignees are to distribute the provided notification to all members of their organization responsible for setting up and supervising patients who use these devices. Additionally, if affected devices were further distributed the notification needs to be provided to them. Philips Respironics is currently investigating the issue and plans to follow-up communications to customers to pro |
| Quantity | 9,999 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of AE, AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, BS, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GF, GI, GR, GT, HK, HR, HU, ID, IE, IL, IN, IT, JE, JO, JP, KE, KR, KW, LB, LT, LY, MA, ME, MK, MM, MQ, MT, MX, MY, NG, NL, |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K181170 | Trilogy Evo Universal Respironics, Inc., Jul 18, 2019 |