Recall Z-0310-2025 ยท MEG Syringe, Antistick

Cardinal Health 200, LLC

Monoject U-100 1 mL Insulin Syringe Luer-Lock with Tip Cap Soft Pack, REF 1188100777

DateNov 8, 2024 (initiated Sep 25, 2024)
StatusOpen, Classified
ReasonCertain lots of listed lots of Cardinal Health Monoject U-100 1 mL Insulin Syringe Luer-Lock with Tip Cap Soft Pack have demonstrated incompatibility with needleless Intravenous (IV) connectors. Cardinal Health recommends that they not be used for administration of IV push insulin via needleless connector.
Root causeUnder Investigation by firm
ActionCardinal Health issued an URGENT MEDICAL DEVICE PRODUCT RECALL to its consignees on 09/25/2024 via overnight mail. The notice explained the issue, risk to health, and requested the following actions: 1. REVIEW inventory for the affected product code and lots. 2. COMMUNICATE with all personnel who utilize the Cardinal Health Monoject U-100 1 mL Insulin Syringe Luer-Lock with Tip Cap Soft Pack that it should not be used for administration of IV push insulin via needleless connector. 3. POST a copy of this notification in your storeroom where the product is stored. 4. NOTIFY any customers to w
Quantity371,326 units
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.