ClearanceBrief
Recall Z-0290-2025 ยท CBK Ventilator, Continuous, Facility Use

Philips Respironics, Inc.

Trilogy Evo O2. Model Numbers: FP2100X10, DS2100X11B, IN2100X15B, DE2100X13B, IA2100X15B, EU2100X15B, LA2100X15B, EE2100X15B, FX2100X15B, ES2100X15B, ND2100X15B, FR2100X14B, IT2100X21B, JP2100X16B, IN2100X19, EU2100X19. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.

DateNov 13, 2024 (initiated Oct 7, 2024)
StatusOpen, Classified
ReasonPotential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.
Root causeDevice Design
ActionAn URGENT Medical Device Correction notification was mailed to consignees on 10/07/2024. The notification provides instructions to users to prevent risks for patients when using in-line nebulizers and a set FiO2. Consignees are to distribute the provided notification to all members of their organization responsible for setting up and supervising patients who use these devices. Additionally, if affected devices were further distributed the notification needs to be provided to them. Philips Respironics is currently investigating the issue and plans to follow-up communications to customers to pro
Quantity24,249 units
DistributionWorldwide distribution - US Nationwide and the countries of AE, AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, BS, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GF, GI, GR, GT, HK, HR, HU, ID, IE, IL, IN, IT, JE, JO, JP, KE, KR, KW, LB, LT, LY, MA, ME, MK, MM, MQ, MT, MX, MY, NG, NL,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181166Trilogy Evo
Respironics, Inc., Jul 18, 2019