ClearanceBrief
Recall Z-0188-2025 ยท DPT Probe, Blood-Flow, Extravascular

Smiths Medical ASD, Inc.

medex TranStar Disposable Transducer 1/EA, Product Code REF MX950

DateOct 23, 2024 (initiated Sep 24, 2024)
StatusOpen, Classified
ReasonDue to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Root causeProcess control
ActionSmiths medical issued an URGENT MEDICAL DEVICE REMOVAL notice to its consignees on 09/24/2024 via traceable mailing. The notice explained the issue, potential risk, and requested the destruction of all affected product. Distributors were directed to forward the notice to their customers.
Quantity1100 units
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.