ClearanceBrief
Recall Z-0216-2025 ยท JEA Table, Surgical With Orthopedic Accessories, Ac-Powered

Baxter Healthcare Corporation

Baxter Operating Table TruSystem 7000 (dV) V, Product code REF 1841083

DateOct 24, 2024 (initiated Sep 19, 2024)
StatusOpen, Classified
ReasonThe firm received customer reports stating that the batteries and their connectors experienced electrical short-circuits and/or emitted smoke.
Root causeDevice Design
ActionBaxter issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 09/19/2024 by USPS first class. The notice explained the issue, the hazard involved, and requested the following: "Actions to be Taken by Customers 1. Once available, Baxter will contact you to implement the design improvement in the impacted surgical tables. Battery replacements must only be performed by personnel authorized, trained, and certified by Baxter. 2. If you received this communication directly from Baxter, please acknowledge receipt of this notification by responding on our customer portal at https:
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.