Recall Z-0181-2025 ยท FET Endoscopic Video Imaging System/Component, Gastroenterology-Urology

Olympus Corporation of the Americas

Olympus Gastrointestinal Videoscope, Model/Catalog Number: GIF-1100

DateOct 23, 2024 (initiated Aug 21, 2024)
StatusOpen, Classified
ReasonIt was discovered during device performance testing that the CCD imaging sensors were programmed with the incorrect color correction data and therefore, specifications are not met.
Root causeUnder Investigation by firm
ActionAn URGENT: MEDICAL DEVICE CORRECTION notification letter dated 9/20/24 was sent to customers. Actions Required: Our records indicate that your facility has purchased one or more of the affected products. Therefore, Olympus requires you to take the following actions: 1. Carefully read the content of this notification. 2. Examine your inventory and identify the above-listed device with the affected serial number. 3. Beginning October 28, 2024, please call Olympus Customer Service at 1-800-848-9024, option 3, to request a CCD Color Adjustment . Olympus will arrange for the return of your device
Quantity1407 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K222584EVIS X1 VIDEO SYSTEM CENTER OLYMPUS CV-1500, COLONOVIDEOSCOPE OLYMPUS CF-HQ1100DL/I, GASTROINTESTINAL VIDEOSCOPE OLYMPUS
Olympus Medical Systems Corporation, Apr 26, 2023