Recall Z-0221-2025 ยท GZL Electrode, Depth

Ad-Tech Medical Instrument Corporation

AD-TECH Spencer Probe Depth Electrode, REF SD08R-AP58X-000 8-Contact 1.12mm (A style) Depth Electrode

DateOct 24, 2024 (initiated Sep 19, 2024)
StatusCompleted
ReasonAd-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol. The A style contacts were not cleared for MR Conditional use. Our records indicate you have received one or more of the affected lots listed above.
Root causeEmployee error
ActionAD-TECH issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 09/19/2024. The notice explained the problem with the devices, risk to health, and requested the consignee contact customer support to obtain an RMA to expedite the return of the affected product. For questions: contact your Ad-Tech Clinical Specialist.
Quantity14 units
DistributionUS Nationwide distribution in the states of CA, OH.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163355Depth Electrodes, Foramen Ovale Depth Eletrodes, Marco-Micro Depth Electrodes, Spencer Probe Depth Electrodes
Ad-Tech Medical Instrument Corporation, Apr 12, 2017