Recall Z-0138-2025 ยท JJW Urinalysis Controls (Assayed And Unassayed)

Mckesson Medical-Surgical Inc. Corporate Office

Quantimetrix Urine Chemistry Control Set Dipper Urinalysis Dipstick Testing Levels 1 & 2 6 X 15 mL-IVD intended as a control for urinalysis reagent strips, microalbumin, and creatinine by listed test methods.

DateOct 22, 2024 (initiated Sep 17, 2024)
StatusOpen, Classified
ReasonThe Quantimetrix Dipper Urinalysis Dipstick control was shipped without the validated summer packaging and may have been exposed to temperatures above 25C. This may lead to delay in patients' results.
Root causePackaging process control
ActionMcKesson issued Urgent Product Recall letter via FedEx Overnight service scheduled for delivery on 9/17/24 to one consignee. Letter states reason for recall, health risk and action to take: 1. Immediately examine your inventory and quarantine all product subject to this urgent product recall detailed in the affected products list (Attachment #1). 2. Please complete, sign, and return the enclosed response form (Attachment #2 - Customer Response Form) as soon as possible, even if you do not have any product on hand. Return completed signed form MMSQARecalls@McKesson.com. 3. If you have any produ
Quantity1 BOX
DistributionUS Nationwide distribution in the state of DC.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K874890QUANTIMETRIX URINE DIPSTICK CONTROL
Quantimetrix Medical Industries, Jan 21, 1988