Recall Z-0228-2025 ยท FNL Bed, Ac-Powered Adjustable Hospital

Baxter Healthcare Corporation

a. Hillrom Centrella Smart+ Medical-Surgical Hospital Bed, Product codes: P7900B000001, P7900B100010, P7900B100024, P7900B100208, P7900B100273, P7900B100304, P7900B000005, P7900B100011, P7900B100025, P7900B100212, P7900B100277, P7900B100307, P7900B000010, P7900B100017, P7900B100026, P7900B100222, P7900B100281, P7900B1SPL05, P7900B000011, P7900B100018, P7900B100028, P7900B100223, P7900B100287, P7

DateOct 25, 2024 (initiated Sep 18, 2024)
StatusOpen, Classified
ReasonThe beds have a potential for an interface issue with the Rauland Responder 5 Nurse Call System where the bed exit system will alert at the bed, but may fail to send a remote alert through the nurse call system.
Root causeSoftware design
ActionThe recalling firm issued letters on and dated 9/18/2024 via first class mail. The letter informed the consignee of the issue, the beds involved, and that Baxter will be upgrading the software on all Centrella Smart+ beds with the affected software version to resolve the issue. An Attachment was enclosed listing the affected product codes for the beds. The hazard involved and the actions to be taken by customers were provided, which included: 1. To mitigate the patient risk associated with this software issue, customers using the Rauland Responder 5 Nurse Call system should ensure the be
Quantity2,967 beds
DistributionWorldwide distribution - US Nationwide. There was government and military distribution. The countries of Canada and Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.