Recall Z-0125-2025 ยท NKB Thoracolumbosacral Pedicle Screw System

XTANT Medical Holdings, Inc

Cortera Spinal Fixation System Screw, Poly, Solid, 6.5mm x 45mm, REF: 150100-65045

DateOct 17, 2024 (initiated Aug 21, 2024)
StatusOpen, Classified
ReasonSpinal fixation system implant screws marked "DNI" are not intended to be implanted and may not have been manufactured in accordance with released manufacturing specifications, and if implanted the screws may not have adequate strength or performance.
Root causeProcess change control
ActionOn 8/21/24, recall notices were mailed and emailed to customers who were asked to do the following: 1) Examine affected devices and quarantine any screws with DNI markings. 2) Notify your staff about the recall. If affected devices were further distributed, provide the recall notice to these customers. 3) Complete and return the response forms via email to mdolan@xtantmedical.com
Quantity330
DistributionUS Nationwide distribution in the states of IL, WI, FL, PA, CA, TX, MI, MO, VA, OH, MA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K221403Cortera Spinal Fixation System
Surgalign Spine Technologies, Aug 19, 2022