ClearanceBrief
Recall Z-0091-2025 ยท QWB Over-The-Counter Molecular Test To Detect Sars-Cov-2 From Clinical Specimens

CUE HEALTH INC

Cue Health COVID-19 Test Cartridge REF 2900005 Cue COVID-19 Test Cue COVID-19 Test for Home and Over The Counter (OTC)

DateOct 15, 2024 (initiated May 24, 2024)
StatusOpen, Classified
ReasonDue to modified test kits being marketed and distributed without FDA Approval, Clearance or Authorization.
Root causeNo Marketing Application
ActionOn 05/13/2024, FDA issued a Field Safety Notice informing/warning the public to not use Cue Health's COVID-19 Tests due to an increased risk of false results. The FDA issued a Warning Letter to Cue Health after an inspection revealed that the company made changes to these tests and that the changes reduced the reliability of the test to detect SARS-CoV-2 virus. Recommendations for Health Care Providers Do not use any Cue Health COVID-19 Tests that you may still have. Dispose of the entire test cartridge in your general waste disposal. Consider retesting your patients using a different FDA
Quantity56 Lots (248,109 total kits)
DistributionU.S. Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.