Recall Z-0005-2025 ยท FRN Pump, Infusion

Zyno Medical LLC

Z-800 Infusion System, Model Numbers Z-800, Z-800F, Z-800W, Z-800WF; Software Version: Z-800 6.1.01 and 6-1.07z; Z-800F 4.1.02 and 4.1.08z; Z-800W 3.1.32 and 3.1.64z, Z-800WF 3.1.32 and 3.1.64z

DateOct 9, 2024 (initiated Sep 13, 2024)
StatusOpen, Classified
ReasonThere is a defect in the air-in-line software algorithm.
Root causeProcess change control
ActionAn "URGENT: MEDICAL DEVICE CORRECTION" notification letter dated 9/13/24 was sent to customers. Mitigation Zyno Medical LLC is mitigating this software issue by conducting a voluntary medical device correction to remove the faulty software and correctly install software that is free of this defect. Recommended Actions 1. Actions to Address the Issue During Use Note that only the devices listed in this letter are affected. No other products are involved in this action. Discontinue use of the affected device and proceed to the instructions for return of the device in Recommended Action b
Quantity34,994 units (1819 units still need correction)
DistributionUS Nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K100705ZYNO Z-800 INFUSION SYSTEM MODEL Z-800
Zyno Medical, LLC, Jul 8, 2010
K130690Z-800 INFUSION SYSTEM
Zyno Medical, LLC, Jun 12, 2013