Recall Z-3167-2024 · OWB Interventional Fluoroscopic X-Ray System

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

System Name: 722044 INTEGRIS Allura 15-12 (biplane) 722030 Integris CV Cesar-Powerpack-Visub-Nicol 722021 INTEGRIS Allura 9 (biplane) 722018 Integris Allura 9 722017 Integris H5000 F / Allura 9F 722016 Integris H5000 C / Allura 9C

DateSep 18, 2024 (initiated Aug 22, 2024)
StatusOpen, Classified
ReasonPotential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.
Root causeDevice Design
ActionPhilips Issued URGENT Medical Device Correction Letter (Philips Reference # 2023-IGT-BST-006) on 8/22/24. Letter states reason for recall, health risk and action to take: a. Affected systems may continue to be used in accordance with their Instructions for Use (IFU) and the instructions below. b. Avoid that operators or bystanders stand underneath the Cable Hose Carriers, especially when moving or turning the system monitor. c. Avoid having the patient located underneath the Cable Hose Carriers (e.g., during transport). d. If you notice loose or broken bolts and/or broken plastic parts of the
Quantity250 units
DistributionNationwide, Canada and Worldwide: Albania Algeria Argentina Armenia Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Brazil Brunei Darussalam Bulgaria Cambodia Cayman Islands Chile China Colombia Costa Rica C¿te D'Ivoire Croatia Cuba Curacao Cyprus Czech
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K002016PHILIPS INTEGRIS ALLURA
Philips Medical Systems North America, Inc., Sep 6, 2000