Recall Z-0086-2025 ยท IDP Microtome, Cryostat
LEICA BIOSYSTEMS NUSSLOCH GMBH
Leica Biosystems Cryostat Model Number CM1510 S, Material Numbers: a) 491510C002, b) 1491510C003; for freezing and sectioning tissue samples
| Date | Oct 15, 2024 (initiated Sep 4, 2024) |
| Status | Open, Classified |
| Reason | As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019. |
| Root cause | Labeling design |
| Action | Leica Biosystems issued an Urgent Field Action Notice to its consignees on 09/05/2024 via mail. The notice explained the problem with the device and requested the placement of a warning sticker regarding the use of flammable freezing sprays on the device and the addition of a supplemental sheet to the current IFU. For questions, please contact: Heather Jensen, Leica Biosystems, 5205 US Highway 12, Richmond, IL 60071 USA, Tel.: +1-815-298-8371, heather.jensen@leicabiosystems.com |
| Quantity | 13 units |
| Distribution | US Nationwide distribution. |
| FDA record | Open on fda.gov |