ClearanceBrief
Recall Z-0064-2025 ยท DXT Injector And Syringe, Angiographic

Merit Medical Systems, Inc.

Blue Diamond Digital Inflation Device, REF: IN7152/C, IN7352/C, IN7403/C, IN7802/C, IN7852/C, IN7130/C

DateOct 11, 2024 (initiated Aug 19, 2024)
StatusOpen, Classified
ReasonInflation devices may have a small hole in the sterile barrier of the device and use of affected devices may result in exposure to pathogens/pyrogens which could lead to infection.
Root causePackaging process control
ActionOn 8/19/24, recall notices were emailed to customers asking them to do the following: 1) If affected devices are discovered, quarantine them, discontinue use and distribution, and return to the recalling firm. 2) Ensure that applicable personnel within your organization are made aware of this field action. 3) If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them and note the quantity distributed on the customer response form. 4) Complete and return the customer response form via email to respo
Quantity14,120
DistributionWorldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and the countries of France, Italy, Jamaic
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K083523MONARCH COMPAK INFLATION SYRINGE AND UNIVERSAL FLUID DISPENSING SYRINGE
Merit Medical Systems, Inc., Feb 4, 2009