ClearanceBrief
Recall Z-0112-2025 ยท DQK Computer, Diagnostic, Programmable

Boston Scientific Corporation

Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO II TO AVVIGO INT UPGRADE JAPAN; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities which includes intravascular ultrasound (IVUS). When used with a peripheral vascular imaging catheter, the system utiliz

DateOct 16, 2024 (initiated Sep 5, 2024)
StatusOpen, Classified
ReasonBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a refere
Root causeUnder Investigation by firm
ActionBoston Scientific issued a Product Advisory notice to its consignees beginning on 09/05/2024 via mail, email or hand delivery. The notice explained the problem with the device, clinical impact, and provided recommendations for safe usage until a software update is available to address the anomaly. Boston Scientific is developing a software update to address this anomaly. The consignees were directed to immediately post this product advisory in a visible location near the AVVIGO+ Multi-Modality Guidance System to ensure this information is easily accessible to users. Also, share this product
Quantity1 unit
DistributionUS Nationwide distribution in Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K230884AVVIGO'+ Multi-Modality Guidance System
Boston Scientific Corporation, Sep 26, 2023