Recall Z-0060-2025 ยท DXT Injector And Syringe, Angiographic

Merit Medical Systems, Inc.

Monarch Inflation Device 30 atm/bar MAP403 MBA Hemostasis Valve Metal Insertion Tool Torque Device REF IN2430 Injector And Syringe, Angiographic/Syringe, Balloon Inflation - The 30 ATM Monarch syringe is used to inflate and deflate balloon angioplasty catheters and to measure the pressure within the balloon during the procedure. It is also used to inject fluids into the body and to monitor the pr

DateOct 11, 2024 (initiated Aug 30, 2024)
StatusOpen, Classified
ReasonDue to a manufacturing process resulting in holes/breaches in the Tyvek sterile barrier.
Root causeNonconforming Material/Component
ActionOn 08/30/2024, the firm email and/or UPS overnight delivery an "URGENT MEDICAL DEVICE RECALL NOTICE" informing customers that specific lots of MONARCH Inflation Devices due to a packaging defect, there may be a small hole in the Tyvek sterile barrier of the device packaging. Use of the affected product may result in exposure to pathogens/pyrogens which could lead to infection. Customer are instructed to: 1. Please immediately determine if any of the devices identified in the attached Customer Response Form (CRF) are within your facility, quarantine them, and discontinue use and distribution.
Quantity6,565 devices
DistributionWorldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Belgium, Canada, France, Ja
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011811MONARCH INFLATION SYRINGE; UNIVERSAL FLUID DISPENSING SYRINGE
Merit Medical Systems, Inc., Aug 22, 2001