Recall Z-0022-2025 ยท JLW Radioimmunoassay, Thyroid-Stimulating Hormone

Qualigen Inc

FastPack TSH Calibrator Kit, REF: 25000024, For use in conjunction with the FastPack IP TSH Immunoassay and FastPack IP System Analyzer

DateOct 8, 2024 (initiated Aug 26, 2024)
StatusOpen, Classified
ReasonThyroid Stimulating Hormone (TSH) calibrator card has an incorrect barcode linked to incorrect information, that may allow successful system calibration, but the TSH results obtained could be higher than expected.
Root causeNonconforming Material/Component
ActionOn 8/26/24, correction notices were emailed to customers who were asked to do the following: 1) Dispose of affected TSH calibration cards. 2) Use TSH Calibration Card Lot Number 2405017-1R sent with correction notices. 3) Run TSH Quality Control samples. 4) Review and retest any affected TSH results that may have been higher than expected. 5) Notify all relevant parties within your organization or any organization that received the affected devices. 6) Complete and return response forms via email to systemsupport@qualigeninc.com If you have any questions regarding this matter, contact Syst
Quantity60
DistributionUS: TX, MS, NY, OH, MT, WA, PA, MN, SC, GA, NC, NV, WY, AK, SD, KS, NE, VT, AL, MA, OK, UT, WI, IA, WV. OUS: Germany, Switzerland
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052301FASTPACK TSH IMMUNOASSAY
Qualigen, Inc., Jan 4, 2006