Recall Z-0020-2025 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Philips North America Llc

IntelliVue Patient Monitor MX400/MX450/MX500/MX550 with software version P.01.01 and ordered with Option CP2 are affected as follows: 1) IntelliVue MX400 866060; 2) IntelliVue MX450 866062; 3) IntelliVue MX500 866064; 4) IntelliVue MX550 866066 Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.

DateOct 4, 2024 (initiated Jun 14, 2023)
StatusOpen, Classified
ReasonDevices with Option CP2 and initially programmed to software version P.01.01 were configured incorrectly at the factory. Therefore, Enhanced ECG Capabilities provided by Option CP2 were not enabled in these devices. Without Option C01 Full Arrhythmia, the device will not provide the yellow alarms for enhanced arrhythmia detection. This is a retrospectively reported recall from 6/14/23.
Root causeProcess control
ActionPHILIPS issued Urgent Field Safety Notice (2023-CC-HPM-019) on 6/14/23. Letter states reason for recall, health risk and action to take: " Please be aware that without Option C01 (Full Arrhythmia) the device will not provide the yellow alarms for enhanced arrhythmia detection. " Review the contents of this letter with your staff. " Pass this notice to all those who need to be aware within your organization or to any organization where the affected devices might have been transferred. 5. Actions taken by Philips to correct the problem Philips Representative will contact you to arrange reload of
Quantity153 units OUS
DistributionForeign Only: Denmark France Germany Japan United Kingdom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K130849INTELLIVUE PATIENT MONITOR MX400; INTELLIVUE PATIENT MONITOR MX450
Philips Medizinsysteme Boeblingen GmbH, Cardiac AN, Jun 14, 2013
K131872INTELLIVUE PATIENT MONITOR MX500, INTELLIVUE PATIENT MONITOR MX550
Philips Medizinsysteme Boeblingen GmbH, Cardiac AN, Oct 17, 2013