Recall Z-0039-2025 ยท LLZ System, Image Processing, Radiological

GE Healthcare

GE Healthcare Centricity Universal Viewer Zero Footprint Client, Model/Catalog/Code Numbers: a) 2089507-120; b) 2089507-129; c) 2089507-131; d) 2089507-135; e) 2089507-136; f) 2089507-140; g) 2089507-144; h) 2089507-145; i) 2089507-149; j) 2089507-153; k) 2089507-154; l) 2089507-155; m) 2089507-156; n) 2089507-157; o) K2037DAED; p) K2037DBED; q) K2037DED; r) K2037DJED; s) K2037

DateOct 9, 2024 (initiated Aug 29, 2024)
StatusOpen, Classified
ReasonGE HealthCare has become aware of an issue in Centricity Universal Viewer Zero Footprint Client (ZFP) version v6.0 SP11.x where the latest report is not shown by default to the user. The issue may occur when a study has multiple DICOM structured reports (i.e. preliminary, final, and/or addendum) stored in different series.
Root causeProcess control
ActionGE HealthCare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 08/29/2024 via mail. The notice explained the issue with the device, safety issue, and provided the following: "Actions to be taken by Customer/User: You can continue to use your device by following the recommendation below. To find the latest report, review all the reports available for the study in the Series Selector. Navigation to the series selection can be found in the User Manual under Section 2.4.3 of The Series Selector, with the subheading of Open a report/document. Please ensure all potential
Quantity1500 units
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131977CENTRICITY UNIVERSAL VIEWER ZERO FOOTPRINT CLIENT
GE Healthcare, Sep 17, 2013