Recall Z-3249-2024 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Philips North America

IntelliVue Patient Monitor MX400 (Model Number 866060), MX430 (Model Number 866061), MX450 (Model Number 866062), MX500 (Model Number 866064), and MX550 (Model Number 866066)

DateSep 25, 2024 (initiated Aug 30, 2024)
StatusOpen, Classified
ReasonDuring a production process, Philips became aware of one IntelliVue power supply with a broken ground bolt upon disconnection of the ground cable from the equipotential ground connector. Loss of electrical grounding may negatively affect the devices electromagnetic immunity and emission.
Root causeNonconforming Material/Component
ActionAn URGENT Medical Device Correction notification letter dated 8/30/24 was sent to customers. Actions that should be taken by the customer / user to prevent risks for patients or users " Please check if the equipotential connector is broken off (refer to Figure 1). If yes, remove the device from use. " Pass this notice to all those who need to be aware within your organization or to any organization where affected devices have been potentially transferred. " Customers should complete the URGENT Medical Device Response Form either via the QR code below or at the end of the notification to submi
Quantity545 devices
DistributionUS Nationwide. Global Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K130849INTELLIVUE PATIENT MONITOR MX400; INTELLIVUE PATIENT MONITOR MX450
Philips Medizinsysteme Boeblingen GmbH, Cardiac AN, Jun 14, 2013
K131872INTELLIVUE PATIENT MONITOR MX500, INTELLIVUE PATIENT MONITOR MX550
Philips Medizinsysteme Boeblingen GmbH, Cardiac AN, Oct 17, 2013