ClearanceBrief
Recall Z-0127-2025 ยท DQK Computer, Diagnostic, Programmable

CHANGE HEALTHCARE CANADA COMPANY

Horizon Cardiology Hemo. Version 12.2.

DateOct 17, 2024 (initiated Aug 21, 2024)
StatusOpen, Classified
ReasonA potential issue has been identified in Change Healthcare Cardiology Hemo where incorrect data entry can lead to inaccurate hemodynamic calculations, potentially resulting in misdiagnosis and inappropriate treatment. This issue occurs when the unit of measurement (UOM) for hemoglobin is configured in a way that differs from the users understanding.
Root causeUse error
ActionChange Healthcare notified consignees on about 08/21/2024 via emailed letter. Consignees were instructed to pay careful attention to the hemoglobin unit of measurement displayed on the top of the "results" screen of the Hemo application, communicate the issue to all relevant team members to ensure vigilance in data entry, ensure that the hemoglobin values are entered in the correct unit of measurement (mmol/l or g/dl) as configured in the Management Console and displayed at the top of the results screen, contact Change Healthcare support to schedule the deployment of the software update tha
Quantity3 systems
DistributionWorldwide distribution - US Nationwide and the countries of Australia, Canada, Germany, Ireland, Israel, United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050561WINDSURFER
Medcon , Ltd., May 4, 2005