Recall Z-0126-2025 ยท MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)
CHANGE HEALTHCARE CANADA COMPANY
Change Healthcare Cardiology Hemo. Versions 14.2, 14.1.1, 14.2, 14.3, 14.3.2, 14.3, 15.0.1, 15.1.
| Date | Oct 17, 2024 (initiated Aug 21, 2024) |
| Status | Open, Classified |
| Reason | A potential issue has been identified in Change Healthcare Cardiology Hemo where incorrect data entry can lead to inaccurate hemodynamic calculations, potentially resulting in misdiagnosis and inappropriate treatment. This issue occurs when the unit of measurement (UOM) for hemoglobin is configured in a way that differs from the users understanding. |
| Root cause | Use error |
| Action | Change Healthcare notified consignees on about 08/21/2024 via emailed letter. Consignees were instructed to pay careful attention to the hemoglobin unit of measurement displayed on the top of the "results" screen of the Hemo application, communicate the issue to all relevant team members to ensure vigilance in data entry, ensure that the hemoglobin values are entered in the correct unit of measurement (mmol/l or g/dl) as configured in the Management Console and displayed at the top of the results screen, contact Change Healthcare support to schedule the deployment of the software update tha |
| Quantity | 221 systems |
| Distribution | Worldwide distribution - US Nationwide and the countries of Australia, Canada, Germany, Ireland, Israel, United Kingdom. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K131497 | MCKESSON CARDIOLOGY HEMO Mckesson Israel , Ltd., Dec 5, 2013 |