Recall Z-3216-2024 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Siemens Medical Solutions USA, Inc.

ACUSON Redwood 1.0 and 2.0 Diagnostic Ultrasound Systems, REF: 11503314, with software

DateSep 24, 2024 (initiated Aug 15, 2024)
StatusOpen, Classified
ReasonIf ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management
Root causeSoftware design
ActionOn 8/15/2024, correction notices were distributed to customers who were asked to do the following: To avoid potential misdiagnosis by an under-represented volume flow measurement value, use the factory default display configuration of l/min (liters per minute). Review of previous examination results ONLY if a clinical assessment reviewed a volume flow measurement(s) with the unit of measurement configured to display a value in milliliters. Firm's Customer Service Engineer will contact you to schedule a facility visit to update the system or inform you of a remote update when the software up
Quantity1990
DistributionUS: MS, WA, TX, NJ, GA, PN, NY, AL, MN, NE, IN, CT, CA, IL, TN, WI, ND, NM, AR, NC, PR, RI, WY, MD, FL, HI, AZ, SC, MO, OK, VA, ME, KS, MI, LA, IA, OH, SD, NH, KY, AL, OR, MA, GU. OUS: Afghanistan, Albania, Algeria, Andorra, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangl
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K210743ACUSON Redwood Diagnostic Ultrasound System
Siemens Medical Solutions USA, Inc., Apr 1, 2021