Recall Z-3155-2024 · PKC Prosthesis, Total Anatomic Shoulder, Uncemented Metaphyseal Humeral Stem With No Diaphyseal Incursion, Semi-Constrained

Tornier, Inc

stryker TORNIER PERFORM HUMERAL SYSTEM - STEMLESS US Anatomic Nucleus, REF DWM003A; anatomic total arthroplasty of the shoulder

DateSep 16, 2024 (initiated Aug 29, 2024)
StatusOpen, Classified
ReasonOne lot of Perform Humeral Nucleus may exceed the specification for bacterial endotoxin load (max 20 EU/device).
Root causeUnder Investigation by firm
ActionStryker issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 08/29/2024 via mail and email. The notice explained the issue, potential risk, and requested the following actions be taken: identify and isolate all affected product to prevent accidental use, and complete and return the response form. Upon receipt, Stryker will contact the consignee and arrange for the return of the product. Distributor were directed to notify their customers. Stryker issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 08/29/2024 via mail and email. The notice explained the i
Quantity5 units
DistributionUS Nationwide distribution in the states of IL, WI, AZ, NV, SD.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K220418Tornier Perform Humeral System – Stemless
Tornier, Inc., Sep 14, 2022