Recall Z-3309-2024 · PAO Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer + Additive, Cemented

Tornier, Inc

Stryker Tornier Perform Humeral System, Reversed Insert, 45%, REF DWP2393, for shoulder arthroplasty

DateSep 27, 2024 (initiated Aug 29, 2024)
StatusOpen, Classified
ReasonDevices from one lot of Perform" Reversed Inserts may contain a locking ring that was assembled upside down which would prevent the humeral insert from securely seating in the stem or spacer.
Root causeUnder Investigation by firm
ActionStryker issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 08/29/2024 via email. The notice explained the issue, potential risk, and requested the following actions be taken: Identify and isolate the affected units, and respond to the notice via email. Upon receipt of the response, Stryker will arrange for the return of the device. Those consignees who further distributed the product (to other sites or customers) were directed to notify those parties. For questions or concerns, contact: memphis.fieldaction@stryker.com.
Quantity38 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K201315PERFORM Humeral System – Stem
Tornier, Inc., Sep 11, 2020