Recall Z-3162-2024 ยท BXB Exerciser, Powered

Btt Health Gmbh

Marodyne-LiV, part number 3045-133, Low intensity vibration exerciser supporting osteoporosis treatment

DateSep 17, 2024 (initiated Aug 1, 2024)
StatusOpen, Classified
ReasonSome power supply units in Austria and Germany had a voltage drop in the millisecond range, which led to a cancellation of the treatment and subsequent restart of the medical device. The treatment is only restarted by a step-off/step-on (step-on for activation).
Root causeComponent design/selection
ActionBtt notified its distributors on 08/01/2024 via email. The notice explained the problem with the device, impact the the user, and were directed to notify their customers. The power supplies will be replaced.
Quantity62 units
DistributionUS Nationwide distribution in the states of Florida.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.