Recall Z-3169-2024 · JIR Indicator Method, Protein Or Albumin (Urinary, Non-Quant.)

Siemens Healthcare Diagnostics, Inc.

Atellica CH Microalbumin_2 (ALB_2)-For in vitro diagnostic use in the quantitative measurement of microalbumin in human urine on ADVIA Chemistry systems/ Atellica CH and Atellica CI Analyzers

DateSep 18, 2024 (initiated Aug 15, 2024)
StatusOpen, Classified
ReasonAtellica CH Microalbumin_2 (ALB_2) and the ADVIA Chemistry Microalbumin_2 (ALB_2) are not meeting the High-Dose Hook Effect claim/Prozone Effect claim as stated in the Instructions for Use (IFU) on the Atellica CH, Atellica CI Analyzers, ADVIA 1800 Chemistry Systems, ADVIA 2400 Chemistry Systems, and ADVIA Chemistry XPT Systems, may result in erroneously depressed microalbumin patient results.
Root causeNonconforming Material/Component
ActionSiemens Healthcare issued Urgent Medical Device Corrections (UMDC) to US consignees ACHC24-05.A.US, ACHC24-05.A.US.CHC on 8/15/24 via FedX and Urgent Field Safety Notices (UFSN)to foreign consignees ACHC24-05.A.OUS, ACHC24-05.A.OUS.CHC to affected customers informing them of the issue and the impact to results. ADVIA Chemistry customers are advised that lot 665399 meets the IFU prozone claim. Customers will be instructed to discontinue use of and discard the impacted reagent lots indicated in the letter. Atellica CH customers will be advised that lots 242194, 242195, and 242321 meet the IFU
Quantity18483 units
DistributionWorldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bosnia Herzeg, Brazil, Canada, Chile, Colombia, Croatia, Cura¿ao, St. Eus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Ho
FDA recordOpen on fda.gov