ClearanceBrief
Recall Z-3081-2024 ยท FMT Warmer, Infant Radiant

Wipro GE Healthcare Private Ltd.

GE Panda iRES Warmer, Model/Catalog Numbers: 1) M1112198-01045518; 2) M1112198-01045579; 3) M1112198-01047148; 4) M1112198-01047271; 5) M1112198-01048731; 6) M1112198-01048740; 7) M1112198-01052386; 8) M1112198-01054697; 9) M1112198-01068942; 10) M1112198-01084972; 11) M1112198-01090848; 12) M1112198-01093457; 13) M1112198-01117766; 14) M1112198-01139893; 15) M1112198-01156745; 16

DateSep 20, 2024 (initiated Aug 9, 2024)
StatusOpen, Classified
ReasonGE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers. There are multiple electrical safety protections included in the product design. However, because the electrical safety testing in manufacturing was incomplete for these devices, there is the potential for leakage current to exceed IEC 60601 limits.
Root causeProcess control
ActionGE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 08/09/2024 by letter disseminated through a traceable means. The notice explained the safety issue and the potential for adverse impact to user or patient. The notice also provided information for the safe operation of the device until GE Healthcare is able to inspect and test the units. A GE Healthcare representative will contact the consignee to arrange for the correction. Questions: Contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Quantity137 units
DistributionUS and Australia, Denmark, France, Germany, Ireland, ITALY, New Zealand, Poland, Portugal, Slovenia, Spain, SWITZERLAND, United Arab Emirates, United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K201628Panda iRes Warmer, Giraffe Warmer
GE Healthcare, Feb 12, 2021