ClearanceBrief
Recall Z-3085-2024 ยท LHI Set, I.V. Fluid Transfer

Baxter Healthcare Corporation

Baxter EXACTAMIX Inlet, Vented, Syringe Inlet, REF H938176, for use in a pharmaceutical compounding device

DateSep 19, 2024 (initiated Aug 20, 2024)
StatusOpen, Classified
ReasonBaxter Healthcare Corporation has received increased customer reports of particulate matter in the Automated Compounding Device Inlets (disposable inlet), Product Codes H938173, H938174, H938175, H938176, used with the ExactaMix and ExactaMix Pro compounders. Particulate matter has been observed within the inlet primary packaging inlet components, including within the sterile fluid path tubing, before use.
Root causeOther
ActionBaxter issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 08/20/2024 via USPS first class mail. The notice explained the issue, hazard involved, and requested the following: "Actions to be Taken by Customers 1. Disseminate this information to anyone who may interact with the ExactaMix and ExactaMix Pro compounders and the products they produce (Pharmacy and Clinical Staff). 2. Pharmacy Staff: Inspect the inlets before use, including the inlet primary packaging, tubing, connectors, and spikes. Perform the inspection in accordance with the enclosed instructions. o If parti
Quantity2,140,046 units
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K002705EXACTA-MIX 2400 COMPOUNDING SYSTEM, MODEL EXACTA-MIX 2400
Baxa Corp., Mar 28, 2001