Recall Z-3156-2024 ยท OLO Orthopedic Stereotaxic Instrument

Mazor Robotics Ltd

Mazor X. Model Number REF TPL0059; stereotaxic orthopedic instrument

DateSep 16, 2024 (initiated Jul 17, 2024)
StatusOpen, Classified
ReasonSoftware anomalies which can result in two separate issues: A lagging screen display during a navigated procedure; or the potential for a mismatch between the navigated screw and the screw information presented on the screen.
Root causeDevice Design
ActionMedtronic/Mazor Robotics, Ltd. issued an URGENT: Medical Device Software Correction to its consignees on 07/17/2024 via 2-day tracked mailing. The notice explained the issue, potential risk, and provided a workaround until the software update is installed. Medtronic will schedule a time to perform a software update that will resolve these anomalies. Medtronic field service engineers will update all Mazor X systems from version 5.1.1 to version 5.1.2 or later. Questions: contact Medtronic Technical Services at 1-888-826-5603 or rs.navtechsupport@medtronic.com .
Quantity94 devices
DistributionUS Nationwide distribution in PR.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K230064Mazor X System (Mazor X Stealth Edition)
Mazor Robotics , Ltd., Apr 7, 2023