ClearanceBrief
Recall Z-2634-2024 ยท OFT Spinal Epidural Anesthesia Kit

MEDLINE INDUSTRIES, LP - Northfield

Medline convenience kits, labeled as: 1) PAIN SERVICES CDS , Pack Number CDS985445B ; 2) PAIN SERVICES CDS , Pack Number CDS985445C ; 3) PAIN TRAY , Pack Number DYNDA1233B ; 4) PAIN TRAY , Pack Number DYNDA2402A ; 5) C ARM BLOCK PACK, Pack Number DYNDA2795; 6) SINGLE DOSE EPI TRAY, Pack Number DYNDA2965; 7) MAMMOGRAPHY TRAY, Pack Number DYNDH1094; 8) FTMC PAIN TRAY, Pack Number DYNDH1239D;

DateAug 27, 2024 (initiated Apr 8, 2024)
StatusOpen, Classified
ReasonMedical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Root causeNonconforming Material/Component
ActionMedline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request these syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other pr
Quantity1,679,067 units in total
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.