ClearanceBrief
Recall Z-3098-2024 ยท DQA Oximeter

Zoe Medical Incorporated

Spacelabs Ultraview DM3 Vital Signs Monitor (all variations -M, -MT, -N, -NT), Model Number 91330

DateSep 10, 2024 (initiated Aug 13, 2024)
StatusOpen, Classified
ReasonThis product labeling does not adequately provide warning about the risks associated with the use of damaged batteries or batteries past the intended use life.
Root causeLabeling design
ActionAn URGENT MEDICAL DEVICE NOTIFICATION letter was sent to customers on 8/13/24. Actions being taken by Spacelabs Healthcare: The Operations Manual and Service Manual for the DM3 and DM4 have been updated to convey the risks associated with continuing to use the lithium batteries past the specified use life and to clarify that batteries older than two years, damaged, or that are no longer holding a charge must be replaced. Actions to be implemented by users: Users must download current versions of the Operations Manual and Service Manual and dispose of any previous copies. The updated product
Quantity6429 units
Distributionawaiting additional details
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K093802ULTRAVIEW DM3 MONITOR
Zoe Medical, Inc., Apr 9, 2010