Recall Z-0003-2025 ยท OZP Automated Insulin Dosing Device System, Single Hormonal Control

Medtronic MiniMed, Inc.

MiniMed Insulin Pump Device. MiniMed 670G, REF: MMT-1580, MMT-1581, MMT-1582, MMT-1780, MMT-1781, MMT-1782, MMT-1740, MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762; MiniMed 770G, REF: MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892; MiniMed 780G, REF: MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896

DateOct 3, 2024 (initiated Jul 31, 2024)
StatusOpen, Classified
ReasonInsulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stoppi
Root causeDevice Design
ActionOn 7/31/24, Safety Alert notices were sent to customers, healthcare professionals, distributors who were told the following: 1) Always pay attention to alerts and alarms displayed on your pump. - Low Battery Pump alert will display when your pump has up to 10 hours of battery life left. - Replace Battery alert will display when the pump has less than 30 minutes of battery life left. If the battery is not replaced within 10 minutes, a siren will sound and repeat once every minute until the Replace Battery Now alarm is displayed. At that time, the pump will stop insulin delivery. If the battery
Quantity526,558
DistributionUS: CT, MI, PA, WA, IA, NY, ND, AZ, TX, OH, NC, AL, MN, IN, NJ, KY, UT, CA, FL, VA, MS, NM, NV, TN, GA, MA, NH, OK, VT, IL, ME, SC, LA, WY, RI, SD, KS, WI, MD, CO, DE, AR, AK, ID, MO, NE, WV, MT, OR, DC, HI, VI, PR. OUS: Worldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.