ClearanceBrief
Recall Z-3172-2024 ยท HTY Pin, Fixation, Smooth

Smith & Nephew, Inc.

DIAMOND POINT, BOTH ENDS K-WIRE, REF 128062, metallic bone fixation fastener

DateSep 19, 2024 (initiated Jul 9, 2024)
StatusOpen, Classified
ReasonSmith & Nephew, Inc., has initiated a field action to voluntarily remove two batches of K-Wires due to a packaging error. A complaint was received indicating that a package contained K-WIRE .062X9 2PT DM wires instead of K-WIRE .045X9 2PT DM as described on the product label.
Root causePackaging change control
ActionSmith & Nephew issued and Urgent Medical Device Recall Notice to its consignees on 07/09/2024 via overnight mail. The notice explained the problem, potential risk, and requested the return of the product. Sales representatives, District Offices, and/or Distributors were directed to notify their customers. For questions or concerns regarding this recall please contact FieldActions@smith-nephew.com.
Quantity73 units
DistributionUS, Canada, Norway
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.