Recall Z-3144-2024 ยท FPA Set, Administration, Intravascular

B. Braun Medical, Inc.

AET36 ANESTH EXT SET - Anesthesia Extension Set - Normally Closed Dual Manifold, Four-way Stopcock, Model/Catalog Number: 472036

DateSep 13, 2024 (initiated Aug 8, 2024)
StatusOpen, Classified
ReasonPotential for product leakage from the Spin Lock Connector on the distal end of the device due to the migration of excess solvent during the assembly process.
Root causeProcess change control
ActionAn URGENT MEDICAL DEVICE RECALL NOTIFICATION letter dated 8/6/24 was sent to customers. Actions Required by B. Braun Medical Inc. (BBMI) Customer/User: 1. Review the Medical Device Recall Notification in its entirety and ensure that all users in your organization of the above-mentioned product, and other concerned persons are informed about this voluntary recall. If you are a distributor and have further distributed the product, please forward this notice to your consignees. The recall is to be extended to the consumer level. 2. Determine your current inventory of the affected items within in
Quantity19050 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022895ELCAM STOPCOCKS AND MANIFOLDS
Medical Device Consultants, Inc., Oct 18, 2002
K111016SAFEPORT MANIFOLD (TM) (OR STOPCOCK)
Elcam Medical Acal, Jun 9, 2011
K943770LOW VOLUME MULTIPORT
B. Braun of America, Inc., Apr 10, 1995