Recall Z-3135-2024 ยท DSP System, Balloon, Intra-Aortic And Control

Datascope Corp.

CARDIOSAVE RESCUE INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate intra-aortic balloons.

DateSep 20, 2024 (initiated Aug 8, 2024)
StatusOpen, Classified
ReasonFirm has developed a software correction (Revision D.00) outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and Cardiosave Rescue Intra-Aortic Balloon Pump devices.
Root causeSoftware design
ActionConsignees were mailed an URGENT FIELD SAFETY NOTICE dated 8/5/2024. This notification informs consignees of the deployment of previously installed software to mitigate previously identified issues in Cardiosave Hybrid and/or Cardiosave Rescue Intra-Aortic Balloon Pump devices. Consignees are asked to examine their inventory immediately to identify if they have any Cardiosave Hybrid and/or Rescue IABP devices, and to identify what software version is currently installed in devices. Devices with software revision D.00 or D.01 have been corrected. If consignee units have a software version lower
Quantity5,475 units
DistributionInternational distribution to the countries of UNITED ARAB EMIRATES, ARGENTINA , AUSTRIA, AUSTRALIA, BELGIUM, BULGARIA, BAHRAIN, BRAZIL, CANADA, SWITZERLAND, CHILE, CHINA, COLOMBIA, CZECH REPUBLIC, GERMANY, DENMARK, ECUADOR, EGYPT, SPAIN, FINLAND, FRANCE, UNITED KINGDOM, INDONESIA, IRELAND, INDIA, I
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181122CARDIOSAVE Intra-Aortic Balloon Pump
Datascope Corp., May 31, 2018