ClearanceBrief
Recall Z-2960-2024 ยท OZD Temporary Non-Roller Type Left Heart Support Blood Pump

Abiomed, Inc.

Impella CP Smart Assist Set APAC; Product Code: 0048-0047;

DateSep 4, 2024 (initiated Aug 5, 2024)
StatusOpen, Classified
ReasonA potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
Root causeUnder Investigation by firm
ActionOn August 5, 2024 URGENT MEDICAL DEVICE CORRECTION (NOTIFICATION) letters were sent to consignees. ACTIONS TO BE TAKEN BY CUSTOMER/USER: " Product is NOT being removed from the field and does not need to be returned. " Review, complete all fields, sign, and return the attached business response form (BRF) (refer to Attachment 2) to Abiomed2860@SEDGWICK.COM. " Forward this notice to anyone in your facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the subject products). " If any of the subject products have been forwarded to another facility, contact tha
Quantity705 units
DistributionWorldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Brunei Darussalam, Canada, Croatia, Czeck Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Panama,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.