Recall Z-3115-2024 ยท NSB Patient Personal Hygiene Kit
MEDLINE INDUSTRIES, LP - Northfield
Medline convenience kits containing semi-rigid suction liners labeled as follows: NICU ADMIT KIT, Pack Number DYKA1014G
| Date | Sep 11, 2024 (initiated Jun 27, 2024) |
| Status | Open, Classified |
| Reason | Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use. |
| Root cause | Process control |
| Action | Medline Industries issued a MEDICAL DEVICE RECALL notice to its consignees on 06/27/2024 via US mail and email. The notice explained the problem with the device, potential risk to the patient, and requested the consignee immediately check their stock for the affected item number and the affected lot numbers and destroy any affected inventory. Distributors were directed to notify their customers of the recall and request the affected products be destroyed. For questions: contact the Recall Department at 866-359-1704 or recalls@medline.com. |
| Quantity | 444 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Canada, Panama. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.