ClearanceBrief
Recall Z-3110-2024 ยท FDE Laparoscopy Kit

MEDLINE INDUSTRIES, LP - Northfield

Medline convenience kits containing semi-rigid suction liners labeled as follows: a) MINOR PROCEDURE, Pack Number CDS981509C; b) GENERAL LAPAROSCOPY CDS, Pack Number CDS981512J; c) OR MAJOR CDS, Pack Number CDS981515D; d) MAJOR-LF, Pack Number DYNJ900709C; e) MAJOR-LF, Pack Number DYNJ900709D; f) GENERAL LAPAROSCOPY-LF, Pack Number DYNJ900723C; g) GENERAL LAPAROSCOPY-LF, Pack Number DYNJ9007

DateSep 11, 2024 (initiated Jun 27, 2024)
StatusOpen, Classified
ReasonDue to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
Root causeProcess control
ActionMedline Industries issued a MEDICAL DEVICE RECALL notice to its consignees on 06/27/2024 via US mail and email. The notice explained the problem with the device, potential risk to the patient, and requested the consignee immediately check their stock for the affected item number and the affected lot numbers and destroy any affected inventory. Distributors were directed to notify their customers of the recall and request the affected products be destroyed. For questions: contact the Recall Department at 866-359-1704 or recalls@medline.com.
Quantity358 units
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Panama.
FDA recordOpen on fda.gov