ClearanceBrief
Recall Z-3107-2024 ยท OEZ Cardiovascular Procedure Kit

MEDLINE INDUSTRIES, LP - Northfield

Medline convenience kits containing semi-rigid suction liners labeled as follows: a) OPEN HEART KIT,ALTA BATES HOSP, Pack Number DYK1002174I; b) ENDO KIT, Pack Number DYKE1558B; c) ENDO KIT, Pack Number DYKE1804A; d) VASCULAR-LF, Pack Number DYNJ900714C; e) MAJOR CARDIOVASC ADULT, Pack Number DYNJ905472B; f) AV FISTULA, Pack Number DYNJ908096A

DateSep 11, 2024 (initiated Jun 27, 2024)
StatusOpen, Classified
ReasonDue to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
Root causeProcess control
ActionMedline Industries issued a MEDICAL DEVICE RECALL notice to its consignees on 06/27/2024 via US mail and email. The notice explained the problem with the device, potential risk to the patient, and requested the consignee immediately check their stock for the affected item number and the affected lot numbers and destroy any affected inventory. Distributors were directed to notify their customers of the recall and request the affected products be destroyed. For questions: contact the Recall Department at 866-359-1704 or recalls@medline.com.
Quantity1,646 units
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Panama.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.