ClearanceBrief
Recall Z-3072-2024 ยท OJH Orthopedic Tray

MEDLINE INDUSTRIES, LP - Northfield

ANESTHESIA TURNOVER KIT, Pack Number DYNJAA0377A

DateSep 6, 2024 (initiated Jun 18, 2024)
StatusOpen, Classified
ReasonA non-sterile Mastisol component, which was not labeled as being sterile or non-sterile, was placed in the non-sterile portion of certain sterile convenience kits and distributed. In the event that the user assumes sterility during its use, there is an increased risk of patient infection.
Root causeProcess change control
ActionMedline Industries issued a Recall notice to its consignees on 06/18/2024 via email. The notice explained the issue, potential risk, and directed the product to be stickered to instruct the used to remove the recalled component from the kit and replace from sterile supply. Distributors or those who have transferred the product were directed to notify those to whom the product was distributed of the recall. For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Quantity315 units
DistributionUS: TX, IL, CA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.