Recall Z-2738-2024 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Agfa N.V.

DR 800. Digital Radiography X-ray System.

DateAug 27, 2024 (initiated Jul 29, 2024)
StatusOpen, Classified
ReasonPotential for the front lever chain of the DR 800 table to fail.
Root causeDevice Design
ActionConsignees were sent an URGENT FIELD SAFETY NOTICE by email, dated 7/29/24. In the notice, consignees are asked to check the correct tightening of the floor plate fixing bolts if a high and persistent noise is heard from the device during the tilting or elevation movement of the tabletop. If the noise persists, the device must be stopped immediately in a horizontal position at the maximum height. Agfa will coordinate with consignees to schedule an inspection of affected devices and to install hardware intended to reinforce equipment before 10/15/24. The recall notification is to be shared with
Quantity141 units
DistributionWorldwide - US Nationwide distribution in the states of CA, FL, IL, IN, KS, LA, MA, MO, NJ, NC, OH, OK, SC, VA, WV, WI; and the countries of AU, BE, BW, CA, CO, CZ, DE, ES, FR, GB, GR, IL, IN, IT, KZ, MX, PL, RS, UZ, ZA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K180589DR 800 with MUSICA Dynamic
Agfa Healthcare N.V., Apr 5, 2018
K183275DR 800 with Tomosynthesis
Agfa N.V., Feb 1, 2019