Recall Z-2721-2024 ยท FPA Set, Administration, Intravascular

Fresenius Kabi USA, LLC

LVP Primary Administration Set, Dual-Inlet, Low-Sorbing, Needle-Free Port, Y-Site (Qty 25), Product Code: SET-0013-25. for intravenous infusion.

DateSep 4, 2024 (initiated Aug 1, 2024)
StatusOpen, Classified
ReasonA defect may cause an uncontrolled medication flow condition which may result in a dose greater than prescribed for the patient (i.e., medication overdose).
Root causeProcess control
ActionOn August 1, 2024, consignees were notified by phone with a follow-up "URGENT - VOLUNATERY RECALL" letter sent August 2, 2024. Required Actions for Users: 1. Discontinue use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2. Inform potential users of the product in your organization of this notification and replace with un-affected product. 3. Complete the attached Customer Reply Form, and return it via the fax number, or email address provided in the letter. Option A: if you have no remaining inventory of the aff
Quantity180 cases/4,500 pieces
DistributionUS Distribution to states of: CO, ID, NJ, UT.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K183311Ivenix Infusion System (IIS)
Ivenix, Inc., Jun 7, 2019