Recall Z-2944-2024 ยท NOU Continuous, Ventilator, Home Use
Philips Respironics, Inc.
Trilogy Evo; Model No./Model Description -- AU2110X15B Trilogy Evo, Australia BL2110X15B Trilogy Evo, Benelux BR2110X18B Trilogy Evo, Brazil CA2110X12B Trilogy Evo, Canada DE2100X13B Trilogy Evo O2, Germany DE2110X13B Trilogy Evo, Germany DS2100X11B Trilogy Evo, O2, USA DS2110X11B Trilogy Evo, USA DS2110X11B Trilogy Evo, USA EE2100X15B Trilogy Evo O2, Eastern Europe EE2110X15B Trilogy Evo, Easte
| Date | Sep 5, 2024 (initiated Jul 16, 2024) |
| Status | Open, Classified |
| Reason | Firm has released a mandatory software update Version 1.05.10.00 and a User Manual addendum for affected ventilators. These updates resolve safety issues identified in earlier recalls. |
| Root cause | Device Design |
| Action | On July 16, 2024, the firm began notifying customers through letters titled "URGENT Medical Device Correction" - UPDATE. Customers were informed of the release of mandatory software update version 1.05.10.00 and a User Manual addendum for the Trilogy Evo, Trilogy Evo O2, Trilogy Evo Universal, and Trilogy EV300 ventilators. This letter outlines the safety issues addressed by this update, previously tracked through recalls RES 91832, 92048, 92515, 93609, and 94263. Issues include "Accuracy of Oxygen Delivery", "Environmental Contamination of Device Sensor", "Battery Depleted or Loss of Power A |
| Quantity | 90,910 US; 56,273 ROW |
| Distribution | Worldwide distribution: US (nationwide) and OUS (foreign) countries of: AU, BD, BE, BG, BO, BR, BS, CH, CL, NL, CN, CO, CZ, DE, DK, FR, SE, GB, GF, GR, GT, HK, HR, HU, IE, IL, IN, IT, JP, KE, KW, LB, LT, MA, ME, MA, MT, MQ, MX, NO, NP, OM, PA, PE, PF, PK, PL, PT, PY, QA, RE, RO, RS, RU, SA, SE, SI, |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K181170 | Trilogy Evo Universal Respironics, Inc., Jul 18, 2019 |