Recall Z-3089-2024 ยท LHN System, Radiation Therapy, Charged-Particle, Medical

Ion Beam Applications S.A.

IBA Proton Therapy System - PROTEUS 235 ProteusPLUS and ProteusONE Software versions: PTS-8 versions, PTS-10 versions, PTS-11 versions, and PTS-12 versions

DateSep 10, 2024 (initiated Jul 8, 2024)
StatusOpen, Classified
ReasonSafety Parameters Verification Mechanisms can be deactivated in clinical runtime
Root causeUnder Investigation by firm
ActionIBA issued AMENDED Urgent Medical Device Correction letter (IBA REFERENCE NCIPT-27531) on 7/10/24 provided by IBA directly to the customer either by hand or by email. Letter states reason for recall, health risk and action to take: Regarding Proteus 235 with a PTS-8, PTS-10, PTS-11 or PTS-12 versions: actions: Regarding the malfunction of Safety Parameters Verification Mechanism within BAPP . IBA carried out an analysis to determine if the impacted sites have some safety parameters checked by the BAPP in their clinical site configuration with values leading to violations. The result of this a
Quantity35 units
DistributionWorldwide - US Nationwide distribution in the states of AK, NJ, PA, VA,OK,FL, KA, IL, LA, MI, TN , TX, VA and the countries of BE, CZ, DE, ES, FR, IN, IT, JP, NL, PL, RU, SE, SG, TW.