ClearanceBrief
Recall Z-2942-2024 ยท NOU Continuous, Ventilator, Home Use

Philips Respironics, Inc.

Trilogy Evo EV300 -- Model Number/ Model Description: BL2200X15B Trilogy Evo, O2, Benelux EV300; BR2200X18B Trilogy Evo, O2, Brazil EV300; CA2200X12B Trilogy Evo, O2, Canada EV300; DE2200X13B Trilogy Evo, O2, Germany EV300; DS2200X11B Trilogy Evo, O2, USA EV300; DS2200X11B Trilogy Evo, O2, USA EV300; EE2200X15B Trilogy Evo, O2, Eastern Europe EV300; ES2200X15B Trilogy Evo, O2, Spain EV300; EU2200X

DateSep 5, 2024 (initiated Jul 16, 2024)
StatusOpen, Classified
ReasonFirm has released a mandatory software update Version 1.05.10.00 and a User Manual addendum for affected ventilators. These updates resolve safety issues identified in earlier recalls.
Root causeDevice Design
ActionOn July 16, 2024, the firm began notifying customers through letters titled "URGENT Medical Device Correction" - UPDATE. Customers were informed of the release of mandatory software update version 1.05.10.00 and a User Manual addendum for the Trilogy Evo, Trilogy Evo O2, Trilogy Evo Universal, and Trilogy EV300 ventilators. This letter outlines the safety issues addressed by this update, previously tracked through recalls RES 91832, 92048, 92515, 93609, and 94263. Issues include "Accuracy of Oxygen Delivery", "Environmental Contamination of Device Sensor", "Battery Depleted or Loss of Power A
Quantity90,910 US; 56,273 ROW
DistributionWorldwide distribution: US (nationwide) and OUS (foreign) countries of: AU, BD, BE, BG, BO, BR, BS, CH, CL, NL, CN, CO, CZ, DE, DK, FR, SE, GB, GF, GR, GT, HK, HR, HU, IE, IL, IN, IT, JP, KE, KW, LB, LT, MA, ME, MA, MT, MQ, MX, NO, NP, OM, PA, PE, PF, PK, PL, PT, PY, QA, RE, RO, RS, RU, SA, SE, SI,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181170Trilogy Evo Universal
Respironics, Inc., Jul 18, 2019